×
近期发现有不法分子冒充我刊与作者联系,借此进行欺诈等不法行为,请广大作者加以鉴别,如遇诈骗行为,请第一时间与我刊编辑部联系确认(《中国物理C》(英文)编辑部电话:010-88235947,010-88236950),并作报警处理。
本刊再次郑重声明:
(1)本刊官方网址为cpc.ihep.ac.cn和https://iopscience.iop.org/journal/1674-1137
(2)本刊采编系统作者中心是投稿的唯一路径,该系统为ScholarOne远程稿件采编系统,仅在本刊投稿网网址(https://mc03.manuscriptcentral.com/cpc)设有登录入口。本刊不接受其他方式的投稿,如打印稿投稿、E-mail信箱投稿等,若以此种方式接收投稿均为假冒。
(3)所有投稿均需经过严格的同行评议、编辑加工后方可发表,本刊不存在所谓的“编辑部内部征稿”。如果有人以“编辑部内部人员”名义帮助作者发稿,并收取发表费用,均为假冒。
                  
《中国物理C》(英文)编辑部
2024年10月30日

An improved method of energy calibration for position-sensitive silicon detectors

Get Citation
Ming-Dao Sun, Tian-Heng Huang, Zhong Liu, Bing Ding, Hua-Bin Yang, Zhi-Yuan Zhang, Jian-Guo Wang, Long Ma, Lin Yu, Yong-Sheng Wang, Zai-Guo Gan and Xiao-Hong Zhou. An improved method of energy calibration for position-sensitive silicon detectors[J]. Chinese Physics C, 2016, 40(4): 046202. doi: 10.1088/1674-1137/40/4/046202
Ming-Dao Sun, Tian-Heng Huang, Zhong Liu, Bing Ding, Hua-Bin Yang, Zhi-Yuan Zhang, Jian-Guo Wang, Long Ma, Lin Yu, Yong-Sheng Wang, Zai-Guo Gan and Xiao-Hong Zhou. An improved method of energy calibration for position-sensitive silicon detectors[J]. Chinese Physics C, 2016, 40(4): 046202.  doi: 10.1088/1674-1137/40/4/046202 shu
Milestone
Received: 2015-06-23
Fund

    Supported by 100 Person Project of the Chinese Academy of Sciences and the National Natural Science Foundation of China(11405224 and 11435014)

Article Metric

Article Views(1536)
PDF Downloads(62)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, [email protected]
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

An improved method of energy calibration for position-sensitive silicon detectors

    Corresponding author: Zhong Liu,
Fund Project:  Supported by 100 Person Project of the Chinese Academy of Sciences and the National Natural Science Foundation of China(11405224 and 11435014)

Abstract: Energy calibration of resistive charge division-based position-sensitive silicon detectors is achieved by parabolic fitting in the traditional method, where the systematic variations of vertex and curvature of the parabola with energy must be considered. In this paper we extend the traditional method in order to correct the fitting function, simplify the procedure of calibration and improve the experimental data quality. Instead of a parabolic function as used in the traditional method, a new function describing the relation of position and energy is introduced. The energy resolution of the 8.088 MeV α decay of 213Rn is determined to be about 87 keV(FWHM), which is better than the result of the traditional method, 104 keV(FWHM). The improved method can be applied to the energy calibration of resistive charge division-based position-sensitive silicon detectors with various performances.

    HTML

Reference (12)

目录

/

DownLoad:  Full-Size Img  PowerPoint
Return
Return